Many GC problems can be prevented if the column is properly installed and GC is maintained routinely. For example, replacing septa or liner at regular intervals and keeping the injector and detector clean and well-maintained shoud solve many problems. Regular preventive maintenance depends on particular model of GC and you should consult required operations in the operator's and service manuals.
For the problem identification we recommend to use electronic flowmeter and leak detector.
Baseline problems could be divided into 5 categories: drift, noise, offset, spiking and wander.
Possible cause | Suggestions |
---|---|
Downward drift for a few minutes is normal after installing a new column | Increase the oven temperature to close to the maximum continuous operating temperature for the column. Maintain the temperature until flat baseline is observed. If the detector signal does not drop in 10 minutes, immediately cool the column and check for leaks. |
Unequilibrated detector | Allow sufficient time for temperature equilibration of the detector. |
Downward drift is frequently due to the "back-out" of contaminants from the detector or other parts of the GC | Clean out contamination. |
Possible cause | Suggestions |
---|---|
Damage to the stationary phase of the GC column | Determine the cause of the damage. It may be due to impurities in the carrier gas or to excessive temperatures. Replace column. |
Drift in gas flow rates | Clean or replace flow or pressure regulator(s). Adjust pressure. |
Possible cause | Suggestions |
---|---|
The column may be inserted too far into the flame of an FID, NPD or FPD detector | Reinstall the column. Be sure to insert the column into the detector exactly the correct distance specified in the instrument manual. |
An air leak can result in noise in ECD and TCD detectors | Eliminate the leak. |
Incorrect combustion gases or flow rates can generate nois in FID, NPD or FPD detectors. | Be sure yuor gases are the proper grade, as well as clean and dry. Reset the flow rates of the gases to their proper values. |
Contaminated injector | Clean injector. Replace inlet liner, septa and selas. |
Contaminated column | Bake out the column. Cut off first 10 cm of column. If it does not help, replace the column. |
Defective detector | Clean and/or replace parts as necessary. |
Defective detector board | Consult GC manufacturer. |
Possible cause | Suggestions |
---|---|
Line voltage changes | Monitor line voltage for correlation with offset. If correlation is found, install voltage regulator or ensure stable power supply. |
Poor electrical changes | Check electrical connections. Tighten any loose connections. Clean any dirty or corroded connections. |
Contaminated injector | Clean injector. Replace inlet liner, septa and selas. |
Contaminated column | Bake out the column. Cut off first 10 cm of column. If it does not help, replace the column. |
Column inserted too fat into the flame of FID, NPD or FPD detectors | Reinstall the column. Be sure to insert the column into the detector exactly the correct distance specified in the instrument manual. |
Contaminated detector | Clean the detector if possible. |
Possible cause | Suggestions |
---|---|
Electrical disturbances entering the chromatograph through power cables, even shielded cables | Try to correlate spikes with events in equipment near the chromatograph. Periodicity is often a clue. Turn off equipment or move it. If necessary, install a voltage regulator. |
Possible cause | Suggestions |
---|---|
Baseline wandering may be caused by changes in environmental conditions such as temperature or line voltage | Try to correlate the wandering with environmental parameters. If a correlation is observed, you will know what to do. |
Inadequate temperature control Check if variations can be correlated with changes in the baseline position. | Measure detector temperature. |
Wandering while using isothermal conditions may be due to contaminated carrier gas | Change the carrier gas or the gas purification traps. |
Contaminated injector | Clean injector. Replace inlet liner, glass wool and seals. |
Contaminated column | Bake out the column. Cut off first 10 cm of column. If it does not help, replace the column. |
Possible cause | Suggestions |
---|---|
Sample validity | Check the concentration and stability of the sample. |
Possible cause | Suggestions |
---|---|
Detector overload. The broad peaks may have a rounded top or even valleys in the top. | Reduce sample volume, dilute with solvent or use higher split ratio. |
Overload of the signal processing electronics. The peaks are clipped with flat tops. | Attenuate detector output or reduce sample amount (see above). |
Fronting peaks are usually the result of column overloading. In this case, the effect should be a function of injection volume. Solutions include reducing the injection volume or using a column with greatr capacity. Columns with larger diameter or thicker stationary phase coatings generally have larger sample capacities, however, resolution may be reduced.
Possible cause | Suggestions |
---|---|
Remmants of previous samples in the inlet or column. Ghost peaks due to remmants are most likely to occur when increasing inlet or column temperatures. | Increase the final temperature and lengthen the run time to allow for complete elution of previous samples. If ghost peaks continue to occur, clean the inlet. Condition the column at temperature higher than has been used but lower than the maximum continuous operation temperature for the column. Cut 10 cm off the inlet end of the column and/orreverese it before reconditioning it. If it does not help, replace the column. |
Backflash may cause remmants. Backflash refers to vapours from the sample which expand to exceed the volume of the injector liner. These vapours may come in contact with colder spots, such as the septum and gas inlets of the injector. Less volatile components may condense. These condensates may vaporize later and interfere with subsequent analyses, sometimes producing "ghost peaks". |
|
Bleed from the septum or fragments of the septum lodged in the inlet or liner. | Clean the inlet. Replace the inlet liner, glass wool and seals. |
Possible cause | Suggestions |
---|---|
Inconsistent injection | Develop a reproducible injection technique. Use autosampler. |
Distorted peak shapes can adversely affect quantitative determinations | Correct any problems that result in the distortion of peak shape. See Peak shape problems. |
Baseline disturbances | See Baseline problems. |
Variations in GC operating parameters | Standardize operating parameters. |
Possible cause | Suggestions |
---|---|
Incorrect polarity of the recorder | Reverse polarity of recorder connections. |
Incorrect setup in the software | Set-up right parameters in your chromatography software. |
Sample compound has greater thermal conductivity than the carrier gas and you are using a TCD or µTCD detector | If possible, change carrier gas. Otherwise there is not a solution. |
Detector overload in element-specific detectors such as ECD, NPD, FPD, etc., can produce both positive and negative peaks | Have the compound of interest arrive at the detector at a different time from the solvent or other compounds in high concentration. H2 produces negative peaks with TCD (µTCD) and helium carrier gas. |
Dirty ECD detector can give negative peak after a positive one | Clean or replace the ECD detector. |
Possible cause | Suggestions |
---|---|
Defective syringe | Try a new or proven syringe. |
"Blown" septum or massive leaks at the inlet | Find and fix leaks. |
Problems with carrier gas flow | Adjust gas flow. Check the column flow ath the column outlet by immersion to methanol. |
Broken column or column installed in the wrong way | Replace or reinstall the column. |
The detector is not functioning or not connected to the recorder or integrator. | Ensure that detector is working properly. E.g.: Is the flame in a FID on? Check connection to the output device. |
Possible cause | Suggestions |
---|---|
Contamination of column and/or liner can lead to loss of sensitivity for active compounds | Clean liner. Bake out the column or replace it. |
Injector leaks reduce the peak height of the most volatile components of a sample more than less volatile | Find and fix any leaks. |
Initial column temperature too high for splitless injection which can prevent refocusing of sample. This affects the more volatile components most. | Initial column temperature should be below the boiling point of the solvent. Decrease the initial column temperature or use less volatile solvent. |
Inlet descrimination. Injector temperature is too low. Later eluting and less volatile compounds have low response. | Increase injection temperature. |
Possible peaks | Suggestions |
---|---|
Fluctuations in column temperature | Repair temperature control system |
Mixed sample solvent for splitless or on-column injections | Use single solvent |
When using injection techniques that require "solvent effect" refocusing such as splitless injectiion, the solvent must form a compact, continuous flooded zone in the column. If the solvent does not wet the stationary phase sufficiently as might be the case for methanol used with a nonpolarstationary phase, the solvent flooded zone may be several meters long and not of uniform thickness. This will result in broad and distorted peaks because the solutes will not be refocused into a narrow band near the beginning of the column. | Installing a retention gap (5 meters of uncoated but deactivated column) ahead of the crhomatographic column may reduce or eliminate this problem. |
Possible cause | Suggestions |
---|---|
Contaminated or active injector liner, seal or column | Clean or replace injector liner. Do not use glass wool in the liner. If necessary, replace the column. |
Dead volume due to poorly installed liner or column. | Confirm by injecting inert peak (methane). If it tails, column is not properly installed. Reinstall liner and column as necessary. |
Ragged column end | Score the tubing lightly with a ceramic scoring wafer or sapphire scriber before breaking it. Examine the end using magnifying glass. If the break is not clean and the end square, cut the column again. Point the end down while breaking it and while installing a nut and ferrule to prevent fragments from entering the column. Reinstall the column. |
A bad match between the polarities of the stationary phase and the solvent | Change the stationaryphase. Usually polar analytes tail on no-polar columns, or dirty columns. |
A cold region in the sample flow path | Remove any cold zones in the flow path |
Debris in the liner or column | Clean or replace the liner. Cut 10 cm off the end of the column and reinstall it. |
Injection takes too long. | Improve injection technique. |
Split ration is too low | Increase split ratio to at least 20:1 |
Overloading the inlet | Decrease the sample volume or dilute sample |
Some types of compounds such as alcoholic amines, primary and secondary amines and carboxylic acids tend to tail. | Try a more polar column. Make a derivative of the dsample. |
Possible cause | Suggestions |
---|---|
Change in column temperature | Check GC oven temperature |
Change in gas flow rate (linear velocity) | Inject a detectable unretained sample such as methane to determine the linear gas velocity. Adjust gas pressure or flow to obtain proper values for your analytical method. |
Leak in the injector | Check the septum first. Change, if necessary. Find the leak and fix it. |
Change of solvent | Use the same solvent for standards and samples. |
Contaminated column | Bake out the column. Cut 10 cm off the end of the column. If necessary, replace the column. |
Possible cause | Suggestions |
---|---|
Damage to stationary phase of column | Replace the column. This is usually indicated by excessive column bleeding or peak tailing. |
Injector problems | Check for leaks, inapropriate temperature, split ration, purge time, dirty liner, glass wool in liner. |
Large increase in sample concentration |
|
Possible cause | Suggestions |
---|---|
Bad column installation | Reinstall column |
Injector leak | Find and fix leak |
Injection volume too large | Decrease sample size or dilute it |
Injection temperature too low | Increase injection temperature so the entire sample is vaporized "instantly". An injection temperature higher than the temperature limit of the column will not damage the column. |
Split ratio is too low | Increase split ratio. |
Column temperature too low | Increase column temperature (ba careful on maximum column temperature limit). Use a lower boiling solvent. |
Initial column temperature too high for splitless injection | Decrease the initial column temperature. Use a less volatile solvent so the initial column temperature is below the solvent boiling point. |
Purge time too long (splitless injection) | Use a shorter purge valve close time. |
Possible cause | Suggestions |
---|---|
Broken column | Replace column. Avoid damaging the polyimide coating on the column. Avoide temperatures above maximu column temperature limit. Avoid abrasion of the column. Remember, even if the column does not break immediately, when protective coating is damaged the column may possibly break spontaneously later. |
Column too hot for too long | Replace the column. Stay below limits specified for the column. |
Exposure to oxygen, particularly at elevated temperatures | Find and fix any lieaks. Be sure carrier gas is sufficiently pure. |
Chemical damage due to inorganic acids or bases | Keep inorganic acids or bases out of column. Neutralize samples. |
Contamination of the column with nonvolatile materials | Prevent nonvolatile materials from getting into column. For expample, use a guard column. |
Detailed information about GC column installation is available here.
In this article, we will show you why GC/MS-TOF is a more suitable technique for modern GC/MS analyzes than the decades-old quadrupole analyzer technology.
The higher speed of the Time of Flight GC/MS system means a higher number of spectra per chromatographic peak. The analyst therefore has more data at his disposal and, thanks to the unique deconvolution algorithm, can identify compounds that have similar properties (isomers, etc.), when resolution by mass spectrum at the unit level is not enough for us.
TOF can work with MS spectrum analysis in the entire m/z range – thanks to good sensitivity, there is no need to use SIM mode and thus the analyst does not lose qualitative information about the structure of the compound. This can play a big role in inter-laboratory control or in cases where there are different results between laboratories.
Time off Flight is not a scanning analyzer, but a pulse analyzer. Thanks to the high ejection rate of 30kHz, it enables the collection of spectra up to 1000/s, which a quadrupole analyzer cannot achieve. In addition, the quadrupole must return to the original value of the electric field after one scan, which takes additional time, the so-called "Interscan time". This significantly increases the average time of one scan. E.g. if PBDE analysis is performed in the range of 100-1000 amu, a GC/MS system with a scanning speed of 20000 amu/s, we will achieve a data collection rate of <20 spectra/s. With the TOF analyzer, we can easily achieve 200 spectra/s, i.e. 10x more. Higher speed enables not only Fast GC/MS, but better quality data. Fast GC/MS provides higher peaks, thus a better signal-to-noise ratio. In addition, in the analysis of compounds such as PBDE, it is necessary to analyze the entire mass spectra, and in the quadrupole analyzer, we thus lose a lot of sensitivity in EI.
When using GC/MS-TOF, we achieve a high speed of collecting spectra, so we can shorten the analysis by 3x to 4x when using a suitable column (0.10 or 0.18 mm ID). This will significantly save on device operating time, carrier gas consumption and increase sample throughput. This can not only increase the capacity of the laboratory, but also speed up the delivery of analysis results.
For most chromatographers poor sample reproducibility and mass transfer onto chromatographic column are generally the most critical issues in the method developement. Among others poor reproducibility and mass transfer can be caused by the degradation of the compounds by the glass inlet liner or by means of insufficient vaporization. Therefore selection of proper GC liner is very important and depends on the injection technique. Here you will find important information helping you with selectionof proper GC liner.
Issues surrounding the GC injector ports
The most commonly used mass spectrometers in gas chromatography are systems based on quadrupole analyzers. Other analyzers used in GC/MS are ion traps. Both analyzers used the same principle. Four poles are connected into the electrical circuit that generates radiofrequency field. Change of this electrical field enables to generate scanning in time. Fragments comming from ion source through ion optics enters the analyzer. The ions separation is accomplished by applying alternating RF frequency and DC voltage to diagonally opposite ends of the quadrupole, which in turn allows aspecific mass fragment to pass through the quadrupole filter.
Picture 1: Scheme of quadrupole analyzer
The simulated counter-current process (Simulated Moving Bed - SMB) was developed in the early 60th by the Universal Oil Products Company. It was mainly applied to industrial-scale separations, like the xylene separation or the fructose-glucose separation.
There is a strong analogy between the SMB and the TMB process. Under the use of a suitable system of adsorbent and eluent, a feed stream is separated into two withdrawal streams containing the pure components of a binary or pseudo-binary mixture.In the SMB process a large column is divided into a finite number of small sections. A withdrawal tube is situated between two of such sections.These tubes are connected in a cyclic mode with the inlets and outlets via a special designed rotary valve (Knauer). An observer sitting on an in- or outlet remarks anapparent moving of solid counter-current the fluid flow at each switching although the solid is fixed in the column. Therefore, the process is called a simulated counter-current one. The SMB is equal to TMB for an infinite numberof columns.
Volume [µl] | Needle length [mm] | Gauge | Needle ID [mm] | Tip | Gastight |
---|---|---|---|---|---|
5 | 50 | 23 | 0.11 | Cone | × |
10 | 80 | 23 | 0.11 | Cone | × |
10 | 80 | 26 | 0.11 | Cone | × |
10 | 50 | 25 | 0.125 | Cone | × |
10 | 80 | 22 | 0.175 | Cone | × |
10 | 50 | 23 | 0.11 | Cone | yes |
10 | 50 | 23 | 0.11 | Cone | × |
10 | 50 | 26 | 0.11 | Cone | × |
Volume [µl] | Needle length [mm] | Gauge | Needle ID [mm] | Tip | Gastight |
---|---|---|---|---|---|
0.5 | 42 | 26 | 0.1 | Cone | × |
0.5 | 42 | 23 | 0.1 | Cone | × |
10 | 42 | 26 | 0.11 | Cone | × |
10 | 42 | 23 | 0.11 | Cone | × |
10 | 42 | 23 | 0.11 | Cone | yes |
Volume [µl] | Needle length [mm] | Gauge | Needle ID [mm] | Tip | Gastight |
---|---|---|---|---|---|
5 | 42 | 26 | 0.11 | Cone | × |
5 | 42 | 23 | 0.11 | Cone | × |
50 | 42 | 23 | 0.24 | Cone | × |
250 | 42 | 23 | 0.24 | Cone | yes |
Volume [µl] | Needle length [mm] | Gauge | Needle ID [mm] | Tip | Gastight |
---|---|---|---|---|---|
5 | 50 | 26 | 0.1 | Cone | × |
5 | 50 | 23 | 0.11 | Cone | × |
5 | 50 | 26 | 0.11 | Cone | × |
10 | 50 | 26 | 0.11 | Cone | yes |
10 | 50 | 26 | 0.15 | Cone | × |
10 | 50 | 26 | 0.11 | Cone | × |
10 | 50 | 23 | 0.11 | Cone | × |
10 | 50 | 22 | 0.175 | Single hole | yes |
10 | 50 | 23 | 0.11 | Cone | × |
25 | 50 | 26 | 0.11 | Cone | yes |
25 | 50 | 26 | 0.15 | Cone | × |
25 | 50 | 23 | 0.24 | Cone | yes |
100 | 50 | 26 | 0.11 | Cone | yes |
100 | 50 | 23 | 0.24 | Cone | yes |
250 | 50 | 26 | 0.25 | Cone | yes |
500 | 50 | 26 | 0.25 | Cone | yes |
1000 | 56 | 26 | 0.15 | Single hole | yes |
1000 | 56 | 23 | 0.15 | Single hole | yes |
2500 | 56 | 26 | 0.15 | Single hole | yes |
2500 | 56 | 23 | 0.15 | Single hole | yes |
Volume [µl] | Needle length [mm] | Gauge | Tip | Gastight |
---|---|---|---|---|
0.5 | 42 | 26 | Cone | × |
0.5 | 42 | 23 | Cone | × |
0.5 | 42 | 23-26 | Cone | × |
1 | 42 | 23 | Cone | × |
5 | 42 | 26 | Cone | × |
5 | 42 | 23 | Cone | × |
5 | 42 | 23-26 | Cone | × |
10 | 42 | 26 | Cone | × |
10 | 42 | 23 | Cone | × |
10 | 42 | 26 | Cone | yes |
10 | 42 | 23 | Cone | yes |
10 | 42 | 23-26 | Cone | × |
10 | 42 | 23-26 | Cone | yes |
25 | 42 | 23 | Cone | × |
50 | 42 | 23 | Cone | × |
100 | 42 | 23 | Cone | × |
250 | 42 | 23 | Cone | × |