Signal to noise ration improvment in GC

GC chromatogramToday's laboratory needs are:

  • Lower detection and quantitation limits (LOD, LQD)
  • Stability improvment in GC and GC/MS systems
  • More inert and stable GC parts (columns, septa, vials, liners, ...)

Lower detection and quantitation limits can be achieved by:

  • Reducing the Noise
  • Increasing the signal

Prior you use the most inert GC instruments and parts, you should optimize the whole system first. A system with lower background will have a better performance. The main bleed/noise generators are:

  • Sample impurities (contamination present in the sample)
  • Impurities in the carrier gas and gas distribution system
  • Septum degradation (GC septa, vial septa)
  • The ferrules
  • Bleed from inserts and glass wool
  • GC Columns (column bleed)
  • Bleed from outside the system (polyimide, fats, cremes, ...)

Note: Leaks, oxydation and moisture will catalyze the processes.

Before the sample enters the GC system

Very often, the sample gets polluted before it is introduced into the GC system. You can avoid it by analyzing the sample as fresh as possible, and avod contact of the solvent with vial cap (careful shaking, etc...). If the septum of vial is penetretad once by the syringe needle, the solvent vapours get contact to the septum material. Softeners in the rubber/silicon are transported in the liquid sample by the solvent vapours. This effect is accelerated with every septum penetration by syringe needle. The effect can be observed already after three injections from one vial.

Also aggressive matrixes can dissolve the vial septum. So be careful on choosing right septa - see manufacturers solvent compatibility charts.

GC septaGC injection port septum

The septa are made of pure polydimethyl siloxane. The amount of PDMS in a septum is 60 times the amount present in the column. Septum is kept at higher temperatures continuously, which causes bleeding. Septum bleed products are cyclic and in D3-D10 range, which appears as ghoast peaks. Septum flush is recommended to minimize this influence. When you choose the proper GC septa, you should be careful on:

  • Septum diameter/shape (depends on the instrument you use)
  • Septum maximum operating temperature (depends on the GC method)
  • Injection lifetime (sometimes depend on the amount of septa layers)

Note: In most cases the operating temperature and septum stability does not depend on its colour.