GC Zero Air Generators The GC Zero Air Generators produce dry and hydrocarbon-free air, using air from oil-free compressors, thus avoiding the need to use traditional bottles that are often complex to change.
The GC Zero Air Generators produce dry and hydrocarbon-free air, using air from oil-free compressors, thus avoiding the need to use traditional bottles that are often complex to change. The ergonomic, compact and silent design makes the GC series generators the ideal solution for zero air supply in the following analytical applications:
GC Zero Air generators use three steps to transform ambient air into analytical grade air. The external oil-free compressor (not included) delivers air through a high efficiency coalescent/particulate prefilter that traps any large particles that may damage the system. The prefilter has an automatic purge system, and traps oil, water and any other particles larger than 5 microns in size. The air leaving the coalescent filter enters a high-temperature platinum catalyser, which through oxidation eliminates hydrocarbon molecules down to < 0.1 ppm. A final high-grade filter is used to remove 99.99% of particulates greater than 0.5 microns in size.
The reduction of hydrocarbons, including methane, and carbon monoxide to < 0.1 ppm decreases the background noise level and gives the baseline much better stability, considerably increasing detector sensitivity and ensuring precise analytical results.
Just connect a suitable oil free compressor to almost immediately have air at the right grade. To save bench space, the unit can be easily installed on the lab wall. No more bottles in the lab and expensive pipelines for air distribution.
The unit only requires connection to a suitable oil free compressor and the mains: the investment can be paid back in less than a year.
Max. zero air flowrate | 3 Nl/min |
Max. zero air pressure | 6.5 bar |
Zero air hydrocarbon content | <0.1 ppm |
Zero air CO content | <0.1 ppm |
Zero air particles removed | 99.99% with size >0.5 micron |
Zero air temperature | Ambient + 15°C |
Max. inlet hydrocarbon content | 100 ppm |
Max. inlet CO content | 50 ppm |
Max. inlet temperature | 40°C |
Inlet pressure range | 4.5–10 bar |
Inlet compressor | Not included |
Pressure drop | <1 bar |
Stabilization time after startup | 40 min |
Inlet/Outlet connection | 1/4" NPT / 1/8" NPT |
Power | 110–120V 60Hz / 220–240V 50Hz |
Dimensions (HxWxD) | 410×550×230mm |
Weight | 9 kg |
Brand: | GC Zero |
Category: | Laboratory instruments and accessories / Chromatography / Gas Generators |
Max. flowrate | 3l/min |
Gas generated | Zero Air |