Model XFM Aluminium Digital Mass Flowmeter

Principle of operation The stream of gas entering the Mass Flow transducer is split by shunting a small portion of the flow through a capillary stainless steel sensor tube. The remainder of the gas flows through the primary flow conduit.

Your price
on inquiry
Availability: ask us
Product description

Product description

Principle of operation

The stream of gas entering the Mass Flow transducer is split by shunting a small portion of the flow through a capillary stainless steel sensor tube. The remainder of the gas flows through the primary flow conduit. The geometry of the primary conduit and the sensor tube are designed to ensure laminar flow in each branch. According to principles of fluid dynamics, the flow rates of a gas in the two laminar flow conduits are proportional to one another. Therefore, the flow rates measured in the sensor tube are directly proportional to the total flow through the transducer. In order to sense the flow in the sensor tube, heat flux is introduced at two sections of the sensor tube by means of precision-wound heater sensor coils. Heat is transferred through the thin wall of the sensor tube to the gas flowing inside. As gas flow takes place, heat is carried by the gas stream from the upstream coil to the downstream coil windings.

The resultant temperature dependent resistance differential is detected by the electronic control circuit. The measured temperature gradient at the sensor windings is linearly proportional to the instantaneous rate of flow taking place. An output signal is generated that is a function of the amount of heat carried by the gases to indicate mass molecular based flow rates.

Parameters

Parameters

Brand: XFM
Category: Measurement and control / Gas measurement / Flowmeters

Design features

  • Supports up to 23 Engineering Units
  • Stores calibration data for up to 10 gases
  • High and low gas flow alarm limits with preset delay interval
  • Two sets of user-programmable electromechanical SPDT relays
  • Internal conversion factors for up to 32 gases
  • Digital interface
  • Automatic zero offset adjustment
  • Self-Diagnostic tests

Technical specifications

Flow Accuracy ±1% of FS at calibration temperature and pressure.
Repeability ±0.25% of full scale.
Flow Temperature Coefficient 0.15% of full scale/ FC or better.
Flow Pressure Coefficient 0.01% of full scale/psi (6.895 kPa) or better.
Maximum Gas Pressure 3447kPa (500psig)
Gas and ambient temperature 0°C to 50°C (0°C to 50°C), dry gases only
Relative gas humidity 70%
Leak integrity max. 1 × 10–9 smL/sec He
Connections 1/4" compression fittings, other types optiona

Use

MediumVzduch
Max. flowrate10l/min

Material

Fitting materialBrass

Operating conditions

DisplayBez displeje
Analogue output0-5 V DC
Digital outputProfibus