Your basket is empty...
Gas | Concetration | Response |
---|---|---|
Hydrogen sulphide (H2S) | ||
CO | 300 ppm | < 1.5 ppm |
SO2 | 5 ppm | cca 1 ppm |
NO | 35 ppm | < 0.7 ppm |
NO2 | 5 ppm | cca 1 ppm |
H2 | 3000 ppm | 0 ppm |
HCN | 10 ppm | 0 ppm |
HCl | 10 ppm | 0 ppm |
NH3 | 50 ppm | 0 ppm |
PH3 | 5 ppm | cca 4 ppm |
CS2 | 100 ppm | 0 ppm |
Methyl sulphide | 100 ppm | 9 ppm |
Ethyl sulphide | 100 ppm | 10 ppm* |
Methyl mercaptan | 5 ppm | cca 2 ppm |
Ethylene | 100 ppm | < 0.2 ppm |
Isobutylene | 100 ppm | 0 ppm |
Toluene | 10000 ppm | 0 ppm* |
Turpentine | 3000 ppm | cca 70 ppm* |
Note: All data above are valid for sensors by RAE Systems. High levels of polar organic compounds including alcohols, ketones, and amines give a negative response.