The AZURA Compact HPLC isocratic educational system is a small complete analytical isocratic HPLC system. Due to its compact dimensions, the system fits onto every laboratory bench. The AZURA Compact HPLC isocratic is based on the multifunctional AZURA element ASM 2.
The AZURA Compact HPLC isocratic educational system is a small complete analytical isocratic HPLC system. Due to its compact dimensions, the system fits onto every laboratory bench. The AZURA Compact HPLC isocratic is based on the multifunctional AZURA element ASM 2.1L and contains two integrated modules and a manual injecton valve, that is fixed to the side of the system. The integrated modules are a pump and a UV/VIS detector with one variable wavelength. A flow cell is included in the package. The HPLC system comes with an eluent tray for safe storage of up to 6 bottles. The HPLC column can be attached anywhere to the system via magnetic clip. A leak sensor and the capillary guidance as well as the optional control unit with touch screen facilitate the safe and user friendly usage. Modern design with completely demountable fronts and exchangeable optionally colored side panels turn the HPLC system into a visual highlight.
//
Brand: | Azura |
Category: | Laboratory instruments and accessories / Chromatography / Liquid chromatographs |
Integrated modules | Pump P 4.1S with 10 ml pump head stainles steel, UV/VIS Detector UVD 2.1S, Manual injection valve, Leak management |
Operation temperature | 10–40°C |
Operation humidity | 10–90%, non-condensing |
Display | 3.5” touch screen |
Dimensions | 361 × 158.2 × 523 mm (W x H x D) |
Power supply | 100 – 240V; 50 – 60 Hz; maximum 100 W |
Digital control | LAN; |
Flow rate | 0.01 – 10 ml/min, 0.01 ml/min increments |
Delivery pressure | max 400 bar |
Flow rate accuracy | ± 1%, measured at 5 – 50% of flow range using ethanol/water 10:90 |
Flow rate precision | < 0.5% RSD |
Wavelength range | 190 – 500 nm |
Detection channels | 1 |
Data acquisition rate | 50 Hz |
Light source | D2 |
Spectral bandwidth | Δλ 13 nm |
Wavelength accuracy | ± 3 nm |
Noise | 30 μAU |
Drift | 300 μAU/h |
Sensitivity | 6 × 10–5 AU at 1 Hz |
Valve head material | stainless steel |
Maximum pressure stability | 300 bar |
Pressure | 400bar |
LC instrument type | HPLC |