Aluminium Mass Flow Controler, model GFC37A (max.flowrate 50l/min air)

Mass flow controller Compact, self-contained GFC mass flow controllers are designed to indicate and control flow rates of gases. The rugged design coupled with instrumentation grade accuracy provides versatile and economical means of flow control. Design features Rigid metallic construction.

Your price
on inquiry
Availability: ask us
Product description

Product description

Mass flow controller

Compact, self-contained GFC mass flow controllers are designed to indicate and control flow rates of gases. The rugged design coupled with instrumentation grade accuracy provides versatile and economical means of flow control.

Design features

  • Rigid metallic construction.
  • Maximum pressure of 1000 psig (70 bars).
  • Leak integrity 1 × 10–9 smL/sec of helium.
  • NIST traceable certification.
  • Built-in tiltable LCD readout.
  • Local or remote setpoint control.
  • 0–5 Vdc and 4–20 mA signals.
  • Circuit protection.
  • TIO Totalizer option.

**CATALOG NUMBER OF MASS FLOW CONTROLLER IS COMPOSED ACCORDING TO YOUR PARAMETERS. ** What the letters ABCDEF means you can see in technical parameters. Contact us, we will be happy to send you a quote.

Parameters

Parameters

Brand: GFC
Category: Measurement and control / Gas measurement / Flow controllers

Technical parameters

Accuracy GFC 17, 37 a 47: ± 1,0% FS.
GFC 57, 67 a 77 (flow range 20–100%): ± 1,5%; optional enhanced accuracy ± 1,0% FS
GFC 57, 67 a 77 (flow range 0–20%): ± 3%; optional enhanced accuracy ± 1,0% FS
CALIBRATIONS Performed at standard conditions [14.7 psia (101.4 kPa) and 70 °F (21.1°C)] unless otherwise requested.
REPEATABILITY ±0.25% of full scale.
RESPONSE TIME Generally 2 seconds to within ±2% of actual flow rate over 25 to 100% of full scale.
TEMPERATURE COEFFICIENT 0.15% of full scale / FC.
PRESSURE COEFFICIENT 0.01% of full scale / psi (0.07 bar).
OPTIMUM GAS PRESSURE 25 psig (1.73 bars).
MAX. GAS PRESSURE 1000 psig (70 bars) maximum GFC 17, 37, 47. 500 psig (34.5 bars) GFC 57, 67, 77.
TURN DOWN RATIO 40:1.
MAX. DIFF. PRESSURE 50 psi for GFC 17/37/57/67 and 77 (3.4 bars), 40 psi for 47 (2.7 bars).
GAS and AMBIENT TEMP 32 °F to 122 °F (0 °C to 50 °C). 14 °F to 122 °F (-10 °C to 50 °C) – Dry gases only.
MATERIALS FLUID CONTACT a. Aluminum models GFC Series: anodized aluminum, 316 stainless steel, brass and Viton O-rings.
b. Stainless steel models GFC17S, 37S, 47S, 57S, 67S and 77S: 316 stainless steel and Viton O-rings.
Optional O-rings: Buna , EPR and Kalrez .
ATTITUDE SENSITIVITY No greater than +15 degree rotation from horizontal to vertical; standard calibration is in horizontal position.
OUTPUT SIGNALS Linear 0–5 Vdc. (1000 ohms min. load impedance); 4–20 mA (0–500 ohms loop resistance) Max noise ±20mV.
COMMAND SIGNALS Analog 0–5 Vdc or 4–20 mA for remote set point mode; NPN compatible purge /valve off.
CONNECTIONS GFC 17: 1/4" compression fittings. Optional: 6mm, 3/8" and 1/8" compression fittings or 1/4" VCR.
GFC 37: 1/4" compression fittings. Optional: 6mm and 3/8" compression fittings or 1/4" VCR.
GFC 47: 3/8" compression fittings.
GFC 57: 3/8" compression fittings.
GFC 67: 1/2" compression fittings.
GFC 77: 3/4" FNPT fittings. Optional: 3/4" compression fittings.
LEAK INTEGRITY 1 × 10–9 smL/sec of helium maximum to the outside environment.
TRANSDUCER INPUT POWER GFC 17, 37 and 47: Universal +12 Vdc to 26 Vdc, 200 mA maximum.
GFC 57, 67 and 77: +12 Vdc, 800 mA; +24 Vdc, 650 mA optional.
CIRCUIT PROTECTION Circuit boards have built-in polarity reversal protection. Resettable fuses provide power input protection.
DISPLAY 3–1/2 digit LCD, 0.5" high characters.
CE COMPLIANT EN 55011 class 1, class B; EN50082–1.

Explanation of the device code

GFC17A-ABCD-EF-GAS-FLOW

  • A = type of seal
  • B = fitting size (connection)
  • C = display (L = including display, N = no display)
  • D = power supply (2 = 12VDC, 4 = 24VDC for GFC 57, 67 and 77, 6 = 12 to 26 VDC for GFC 17, 47 and 47)
  • E = output (A = 0 to 5 VDC, B = 4 to 20 mA – local, C = 0 to 5 VDC / 0 to 5 VDC, D = 0 to 5 VDC / 4 to 20 mA, E = 4 to 20 mA / 4 up to 20mA, F = 4 to 20mA / 0 to 5 VDC)
  • F = digital output (always = 0)

Use

MediumPlyn
Max. flowrate50l/min

Operating conditions

Analogue outputna výběr
Product inquiry
Visited (1)