0 Porovnat
Přidejte si do porovnání produkty pomocí ikonky vah a zde si poté můžete porovnat jejich parametry.
Uživatel
0 Košík
Váš košík je prázdný...

Křížové interference senzoru oxidu uhelnatého

Plyn Koncetrace Odezva bez filtru# Odezva s filtrem*
Oxid uhelnatý (CO)
H2S 10 ppm 0 ppm
SO2 5 ppm 0 ppm
Cl2 10 ppm 0 - 1 ppm
HCl 10 ppm 0 ppm
NO 35 ppm 1 ppm 1 ppm
NO2 5 ppm 0 ppm 0 ppm
NH3 100 ppm 0 ppm 0 ppm
H2 100 ppm 40 ppm 40 ppm
Etylen 100 ppm 24 ppm 20 ppm
Acetylen 250 ppm 250 ppm
Etanol 200 ppm 0 ppm 0 ppm
Etylene oxid 125 ppm > 40 ppm
Propan 100 ppm 0 ppm 0 ppm
Butan 100 ppm 1 ppm 1 ppm
Isobutylen 100 ppm 9 ppm 4 ppm
Isobutylene 1000 ppm 30 ppm 22 ppm
Hexane 500 ppm 0 ppm 0 ppm
MEK 100 ppm 0 ppm 0 ppm
TCE 100 ppm 25 ppm 15 ppm
Nitrogen 100 % 0 - 4 ppm

Note: All data above are valid for sensors by RAE Systems.

#New sensor. Used sensors show increasing response to VOCs.

*Disk-shaped activated carbon fiber filters placed on top of the CO sensor help reduce response to VOCs.